The HTSY® single syringe Angiographic Injector is a high-pressure infusion device specifically developed by Tianjin HanTeng Shengye Technology Co., Ltd. for use in CT enhanced scanning. The entire machine has a compact structure and is flexible to deploy, suitable for conventional CT rooms, health check centers, and mobile CT diagnosis and treatment scenarios. The equipment is equipped with a complete intelligent safety protection mechanism, ensuring stable injection output and stable image quality for enhanced scanning; it also has the advantages of moderate procurement threshold, strong consumable versatility, and simple operation and maintenance. It is a cost-effective main equipment for CT enhanced examinations in hospitals and clinics at all levels. Interested customers are welcome to inquire about cooperation.
The HTSY® single syringe Angiographic Injector is specially developed for CT enhanced imaging examinations. The device is equipped with a standardized synchronous triggering interface, which can be linked with the CT scanner host to achieve precise timing matching between injection and scanning. Medical staff can freely set the injection flow rate, total dosage, maximum pressure, and delay scanning time. The system supports storing multiple sets of standardized enhanced contrast imaging protocols and supports one-click retrieval and use. The device is compatible with market-standard disposable high-pressure syringes, which can automatically complete drug suction and pipeline exhaust operations; it supports two modes of local panel operation and remote control of compartments, reducing the radiation exposure of medical staff. It is widely used in head, chest, abdomen, limbs, and cardiovascular CT enhanced scans, evenly and stably pushing the contrast agent to obtain clear and well-contrasted sectional images, assisting doctors in completing lesion screening and qualitative diagnosis.



| Injector Tube Capacity | 200ml |
| Injector Capacity Range | 1—200ml |
| Injection Speed Rate | 0.1-10ml/s step |
| Injection Delay | 0—600s |
| Scan Delay | 0--600s |
| Memory Injection Program | 100 sets |
| Injection Change Capacity | 1—8phases available for selection |
| Preset Pressure Limit | 100、150、200、250、300psi |
| Sucking and Exhaust Rate Adjustment | 0.4ml/s 、6ml/s adjustable |
| Power Supply | 220VAC±10% /50HZ±1HZ |
The HTSY® single syringe Angiographic Injector is optimized for CT enhanced scanning scenarios, eliminating the redundant structure of the dual-syringe model. The machine body occupies less space and can easily adapt to the compact standard CT rooms. The equipment is focused on routine enhanced examinations, has a lower procurement cost, and is suitable for batch configuration by medical institutions such as health centers, hospitals, and clinics that mainly conduct CT enhanced business and have moderate contrast agent flow requirements.
This high pressure injector is equipped with a real-time pressure closed-loop monitoring system that dynamically controls the injection pressure and flow rate throughout the process, outputting a smooth contrast agent infusion curve, avoiding sudden changes in flow rate that cause contrast agent shock and extravasation. It supports multi-stage segmented injection parameter settings, adapting to the requirements of multi-phase enhanced scanning in the arterial, venous, and delayed periods, effectively reducing image artifacts and improving the detection rate of lesions.
The HTSY® single syringe Angiographic Injector is equipped with high-precision air bubble detection, pressure over-limit automatic protection, and one-click emergency stop functions; it integrates syringe in-place identification and equipment tilt prevention interlock. When the system detects pipeline air bubbles, pressure abnormalities, installation not in place, etc., it immediately stops the injection and gives an audible and visual alarm, effectively avoiding clinical risks such as air embolism and contrast agent leakage, and ensuring the safety of enhanced scanning patients.
This high pressure injector is equipped with a large-sized color touch screen, with an intuitive operation interface and built-in standardized enhanced solutions for common parts such as chest, abdomen, and head. The pre-drawing, exhaust, and pre-injection procedures are highly simplified, significantly reducing the preparation time for a single patient and improving the average daily examination and reception volume of the CT room, alleviating the workload pressure of the department during peak hours.
The HTSY® single syringe Angiographic Injector is compatible with mainstream specifications of disposable high-pressure syringes and connection pipelines available on the market, without exclusive dedicated consumables binding. Hospitals have more options for consumable procurement. The entire machine shell is made of materials resistant to various medical disinfectants, with a smooth surface without cleaning dead corners; the internal transmission mechanism has been verified for durability, with fewer fault points, simple daily maintenance, and lower long-term operation and maintenance costs.
A general synchronous trigger signal port is reserved, which can be connected to various brands of spiral CT domestically and internationally, achieving synchronous triggering of scanning and contrast agent injection. An optional compartment remote controller is available, supporting medical staff to remotely start and stop the equipment in the operation room, reducing continuous ionizing radiation exposure and improving occupational protection conditions.